Automatic Lap Tester
Model:GW-901RH
Working Power Supply:AC220V±10%, 50Hz±2Hz
Operational Environment Environmental Temperature: +5℃~+35℃, Relative Humidity: 85% below
Overall Dimension:370 x 220 x 110 (mm)
Weight Approx.:7kg
Description
NAME:Automatic Lap Tester (GW-901RH)
Specifications of Sensor | Φ2mm | Φ3mm | Φ4mm | Φ6mm | Φ10mm | |
Measurement Accuracy | 0~300 coils | ±0 coils | ||||
301~500 coils | ±1 coils | |||||
501~5000 coils | ±0.3% | ±0.3% | ±0.2% | ±0.2% | ±0.2% | |
5001~20000 coils | -- | ±0.4% | ±0.2% | ±0.2% | ±0.2% | |
20001~60000 coils | -- | -- | ±0.5% | ±0.5% | ±0.5% | |
Measuring Range | 0~5000 coils | 0~20000 | 0~60000 coils | |||
Inner Diameter of Measured Coil | >2mm | >3mm | >4mm | >6mm | >10mm | |
Height of Measured Coil | ≤20mm | ≤30mm | ≤110mm | |||
External Diameter of Measured Coil | ≤15mm | ≤25mm | ≤120mm | |||
Working Power Supply | AC220V±10%, 50Hz±2Hz | |||||
Operational Environment | Environmental Temperature: +5℃~+35℃, Relative Humidity: 85% below | |||||
Overall Dimension | 370 x 220 x 110 (mm) | |||||
Weight | Approx. 7kg | |||||
No display or only one Nixie tube lights
| 1. Check if the power socket and the fuse are in normal |
2. Check according to each group of voltage marked on the power socket bracket, if not, repair the power supply | |
3. Check pins (30) on main board AT89S52, if there is no high frequency impulse, replace the crystal oscillator 20MHz at main board AT89S52. | |
4. Check if the pin between the display panel and the main board is plugged well. | |
Display instability | 1. The sensor shall be plugged properly with good contact. For discrete type, check if the two connecting wires are broken. |
2. Check whether the sensor works well | |
3. Check if the steel tube of the sensor is connected well with the machine grounding wire or the enclosure | |
Breaking light constantly on | 1. Check if the output clamp is broken |
2. Check if the printed board of the heat sink is broken | |
3. Check if the machine plug-ins are loose or jumped out | |
4. If the discrete connecting wire is disconnected. | |
Shell-touching lamp constantly on | 1. Check if the clamp wire touches the enclosure or the stainless steel tube |
2. Check if the printed board of the heat sink is broken | |
3. Check is the resistance 5.1Ω-0.5W of enlarge board is broken. | |
Big error | 1. If the sensor is matched with the machine |
2. Whether the sensor works well | |
3. If related measuring conditions comply with other precautions |